MA11-EA4313A
MA11-EA4313A  MA11-EA4313A  MA11-EA4313A 

MA11-EA4313A

  • Brand: Huying Intelligen
  • Product Code: MA11-EA4313A
  • Reward Points: 180
  • Availability: In Stock
  • $600.00

  • Ex Tax: $600.00

Qty

Product Introduction
MA11-EA4313A The oxygen content analyzer is the latest high-precision online oxygen content detection and analysis instrument developed by our company. This instrument adopts imported zirconia sensors and combines microcontroller control technology. It has the advantages of advanced technology, high precision, fast response, stable performance, complete functions, convenient operation, and continuous gas analysis process. Widely used in industries such as air separation for nitrogen production, petrochemicals, metallurgy, electronics and power, and healthcare that require detection and analysis of oxygen content in various mixed gases.

Technical parameter

operational principle

Zirconia

Display method

4.3 inch touch screen

Detection scope

0~10/100/1000ppm/1.00/21.00% O2

 

measurement accuracy

0~1000ppm/1.00/21.00%O2≤±1% FS

0 ~ 100ppm≤±2% FS

0 ~ 10ppm≤±5% FS

resolving power

0.01% O2

Repeatability

±1% FS

 

Control output

Non isolated 4~20mA  DC

Non isolated 0~10V    DC

2 Programmable relay alarm output

Communication mode

RS232,RS485,Ethernet

Working power supply

AC85 ~ 264V 50/60Hz

Environmental temperature and humidity

-5℃~+50℃   ,   <80% RH

Sample gas temperature

-10℃ ~ +50℃

Adopting methods

Aspirated type

Sample gas flow rate

2~3L/min

Sample gas pressure

Micro positive pressure, micro negative pressure, or positive pressure

Background gas

A mixture of nitrogen, inert gas, and oxygen

Specifications and dimensions

127mm×271mm×296mm(H×W×D)

service life

>4 years (under normal usage conditions)

Pneumatic interface

Stainless steel quick tightening with a diameter of 4mm inside and 6mm outside

Installation method

Placed or embedded

 

Product dimension diagram

Front view

Rear view

Communication ports and wiring terminals
RS232communication interface
Used to connect external communication devices, the local output interface is a standard DB9 female head.

RS485 communication interface
Used to connect external communication devices, the local output interface is a screw type terminal block.

Definition of wiring terminals

Product features

1、The instrument adopts a 4.3-inch touch screen display, which is easy to operate and displays clearly;
2、The instrument is equipped with RS-232, RS485 communication ports, and an Ethernet interface that complies with the Modbus TCP protocol. Multiple communication ports enable users to easily access the Internet of Things;
3、The instrument can communicate one-way or two-way with the upper computer.
4、Original imported zirconia sensor with high measurement accuracy, fast response, and long service life;
5、Calibration is convenient and simple. It can be calibrated with air or standard gas at a single point, which can meet the precise measurement of oxygen content in the range of ppm~%;
6、Built in temperature and pressure monitoring modules and pipeline pressure sensors;
7、The instrument is equipped with an original imported suction pump, which works stably and has a long service life;
8、Equipped with gas path blockage alarm and self-protection function. After the air path is blocked, the air pump will automatically shut down, stopping the operation of the oxygen sensor, which can effectively extend the service life of the air pump and sensor;
9、Wide range AC power supply, with a wider range of applications;

Be the first to write a review for this product.

Write a review

Please login or register to review